Journal of Liquid Chromatography & Related Technologies · 2001 · 18 citations · 12 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistrySeparation ProcessesSeparation ScienceAnalytical ChromatographyChromatographic SeparationChromatographyLiquid ChromatographyContinuous OperationMultidimensional ChromatographyPolystyreneTwo-dimensional Liquid Chromatography
Abstract The two dimensional separation of low molecular weight polystyrene oligomers and stereoisomers was achieved using a liquid chromatographic system designed for ‘heart-cutting’ multidimensional chromatography. The first separation process incorporated a C18 column and a methanol mobile phase to separate the oligomers, whilst the second separation process utilised the change in selectivity associated with a carbon clad zirconia stationary phase and an acetonitrile mobile phase for the separation of the isomers. The two dimensional separation that was achieved using this automated system was comparable to the individual separations following manual sample manipulation in each individual separation step. Furthermore, the multidimensional LC system was shown to be highly reproducible in routine continuous operation over an eight hour period. ACKNOWLEDGMENTS This work was supported by a UWSH internal research award (3202 7209). APS would like to acknowledge the receipt of an Australian Postgraduate Award (APA).
12
Informational Orthogonality of Two-Dimensional Chromatographic Separations
Patrick J. Slonecker, Xiaodong Li, Thomas H. Ridgway et al. · Analytical Chemistry · 1996 · 173 citations
Systematic approach to column switching
Klaus Ramsteiner · Journal of Chromatography A · 1988 · 73 citations