IEEE Transactions on Instrumentation and Measurement · 1985 · 44 citations · 7 references
EngineeringElectrical EngineeringAnalog ElectronicsCircuit TheoryMicroelectronicsCurrent Measurement InstrumentsCurrent-voltage CharacteristicsOhm's LawPower DividersResistance MeasurementCryogenic Readout ElectronicsDirect Current Resistance MeasurementEquivalent RatiosEquivalent Series ResistancePrecision Resistance MetrologyResistance Standard CalibrationResistivity MeasurementResistors
The cryogenic current comparator (CCC) built at LCIE allows measurement of resistance ratios equal to α(1 ± 4 × 10-5) where α = 1, 10, 100, 64.53, or 129.06. The equipment is now currently used for intercomparison of 1-Ω standard resistors with a typical standard deviation of 5 × 10-9 (one single observation) and for calibration of 10 000-Ω resistors in terms of 1-Ω resistors with a standard deviation of 2 × 10-8 (1 single observation). Comparison of the 100/1 (or 10/1 × 10/1 ratio of the CCC with the ratios of Hamon-type series parallel dividers is reported. The agreement between these different ratios is better than 1 × 10-8.
7