Review of Scientific Instruments · 1958 · 10 citations · 3 references
Natural SciencesPhysicsOpticsRadiation MeasurementSpectroscopyCalibrationOptical SpectroscopyRadiation InstrumentationSpectral CalibrationBlackbody CalibrationBlackbody RadiationCalibration SourcesEnergy CalibrationGrating SpectroscopySpectrometer CalibrationSpectroradiometric Calibration
A blackbody was constructed of graphite which could be inductively heated in an atmosphere of helium to a maximum temperature of 2700°K. This blackbody was used as a source of radiant energy in the calibration of a grating spectroradiometer. The calibration was made over the spectral range of 2600 to 3800 A in the second order, and of 3300 to 7550 A in the first order. The accuracy of the calibrations was estimated to be of the order of 9% in the regions 2600 to 2800 A in the second order, and 7000 to 7550 A in the first order, and of 5% in the regions 2800 to 3800 A in the second order, and 3300 to 7000 A in the first order.
3
A Small Plane Grating Monochromator
W. G. Fastie · Journal of the Optical Society of America · 1952 · 263 citations
Natural Sciences, Physics, Optics +16
Evaluation of the quality of a blackbody
J.C. De Vos · Physica · 1954 · 167 citations