Physics in Medicine and Biology · 1969 · 21 citations · 3 references
EngineeringMaterials EngineeringMaterials ScienceElectronic MaterialsHeat TransferRadiation DosimetryOptically Stimulated Luminescence DosimetryTemperature MeasurementThermal History AnalysisThermal Properties Of MaterialsThermal Properties Of SemiconductorsThermoluminescent DosimetryAnnealing Temperature EffectsCooling Rate EffectsEquilibriumPrecisionReadout InstrumentationThermal Equilibrium MeasurementThermoluminescent DosimetersThermoluminescent Materials
The influence of the thermal history on the yield of thermoluminescence has been investigated. Dosimeters (mostly LiF-Teflon) have been stored at various temperatures for various times (pre-annealing) prior to irradiation and readout. Temperatures as high as 300 degrees C may give the LiF-crystal a state of thermal equilibrium in seconds, while the same equilibrium will not be reached for months at 80 degrees C. For precision dosimetry, pre-annealing at a high temperature should last only long enough to give thermal equilibrium and to empty all filled traps. Especially important is reproducibility of the cooling rate. With an appropriate heating circuit, the demands of both pre-annealing and cooling rate can be satisfied with the readout heating.
3
Thermoluminescent Radiation Dosimetry Utilizing LiF
John Cameron, Dean W. Zimmerman, G.N. Kenney et al. · Health Physics · 1964 · 158 citations