Thermal Annealing Effects on the Thermoluminescence of LiF

Dean W. Zimmerman, Charles R. Rhyner, John Cameron

Health Physics · 1966 · 126 citations · 0 references

Concepts

Abstract

Isothermal annealing before or after irradiation is found to have a definite effect upon the thermoluminescent glow curve of LiF (TLD-100). Post-irradiation annealing, by thermally releasing electrons from their traps, reduces the size of each peak in the glow curve. The two main dosimetry peaks each have a simple exponential decay consistent with the model of Randall and Wilkins, indicating trap depths of 1.19 and 1.25 eV respectively with frequency factors between 1011 and 1012 sec−1. Pre-irradiation annealing at 80°C for 24 hr reduces the area of the low temperature peaks by a factor of 10, while annealing at 125°C, 145°C or 175°C for 3 hr reduces the area of the main dosimetry peaks by a factor of 2–5 and increases the height of one of the low temperature peaks by a factor of 12-3. The rate of cooling from temperatures greater than about 150°C affects the glow curve also. Rapid cooling in times of the order of seconds increases the heights of the low temperature peaks up to a factor of 4 compared to cooling in several minutes or more.