Wavelength Calibrations in the Infrared IV Use of a 1000-Lines-per-Inch Bausch and Lomb Plane Replica Grating

K. Narahari Rao, W. W. Brim, V. L. Sinnett, R. H. Wilson

Journal of the Optical Society of America · 1962 · 25 citations · 5 references

Concepts

Abstract

It has been demonstrated that the 1000-lines-per-inch Bausch and Lomb plane replica grating can be employed for high-resolution infrared studies anywhere in the region between 1 and 40 μ. In this manner, calibration procedures become simplified in measuring wavelengths relative to the CO standards. A map of the pure rotational lines of the H2O molecule along with the vacuum wave numbers of these lines has been presented.

References

5