Publication | Closed Access
Measurement of Stark broadening and shift parameters of several Ar I lines
24
Citations
22
References
1998
Year
Ar IiEngineeringMeasurementOptical TestingInterferometryPlasma SciencePlasma PhysicsStark BroadeningSpace Plasma PhysicShift ParametersPulsed ArcInstrumentationPlasma DiagnosticsPhysicsAtomic PhysicsSynchrotron RadiationStark ShiftsSeveral ArNatural SciencesSpectroscopy
This work reports Stark shifts and widths of 15 Ar I lines measured in a pulsed arc. Plasma parameters, electron density and temperature, range in this experiment from 0.25 to and from 13 500 to 26 500 K, respectively. The former has been measured interferometrically at a single wavelength and spectroscopically from the -Stark broadening. The latter has been measured from the Boltzmann plot of Ar II lines and from the ratio of Ar II and Ar I line intensities. Comparisons with most of the data published to date about this topic in Ar I are also included.