Physics in Medicine and Biology · 1971 · 118 citations · 5 references
Materials ScienceEngineeringEffective Atomic NumberKerma ValuesMass SpectrometryDosimetry MaterialsLithium FluorideSoft TissueRadioanalytical ChemistryLithium BorateAtomic AbsorptionBiomedical EngineeringChemistrySorel CementMedicineNuclear MedicineDosimetryBiophysics
The calculated value of the effective atomic number (ztau ) of 7.23 for lithium tetraborate (Li2B4O7) agrees closely with the value of 7.22 for soft tissue of composition C5H40O18N. For excised muscle specimens the experimental value (7.42) obtained by Spiers for ztau is very nearly equal to the corresponding calculated value (7.46) for the composition (ICRU) of striated muscle. A similar comparison shows that the composition for bone given by Woodard with a higher content of calcium and phosphorus gave a value of ztau =13.3, which is in closer agreement with the measured value of 13.8 obtained by Spiers. The calculated values of ztau and kerma per roentgen for lithium tetraborate and lithium fluoride for photon energies below 1 MeV show that lithium borate (Li2B4O7) is a better substitute for soft tissue than lithium fluoride.
5
<i>Alpha-, Beta-, and Gamma-Ray Spectroscopy</i>
Kai Siegbahn, Evans Hayward · Physics Today · 1965 · 756 citations
D. E. A. Jones, H. C. Raine · British Journal of Radiology · 1949 · 114 citations
Materials for Depth Dose Measurement
F. W. Spiers · British Journal of Radiology · 1943 · 54 citations
Radiation Dosimetry and Protection
L. D. Marinelli · Annual Review of Nuclear Science · 1953 · 29 citations · Full text
A New Tissue-Like Material for Use as Bolus
Duncan D. Lindsay, Babette E. Stern · Radiology · 1953 · 14 citations