Preliminary performance measurements for a streak camera with a large-format direct-coupled charge-coupled device readout

R. A. Lerche, J. W. McDonald, R. L. Griffith, G. Vergel de Dios, D. S. Andrews, A. W. Huey, P. M. Bell, O. L. Landen, P. A. Jaanimagi, R. Boni

Review of Scientific Instruments · 2004 · 16 citations · 1 references

Concepts

Abstract

The University of Rochester’s Laboratory for Laser Energetics (Rochester, New York) is leading an effort to develop a modern, fully automated streak camera. Characterization of a prototype camera shows spatial resolution better than 20 lp/mm, temporal resolution of 12 ps, line-spread function of 40 μm (full width at half maximum) contrast transfer ratio of 60% at 10 lp/mm, system gain of 101 charge-coupled device electrons per photoelectron, and a dynamic range of 500 for a 2 ns window.

References

1