Arabian Journal of Chemistry · 2010 · 20 citations · 3 references
PharmacyChemistryPharmacologyAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisAnalytical ChromatographyBioanalytical Method ValidationChromatographyIonization TechniquesLiquid ChromatographyMass SpectrometryConcentrationElectrospray IonizationLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method DevelopmentLower Limit Of QuantificationLower Limit Of QuantitationTandem Mass Spectrometry
A sensitive and selective liquid chromatographic–tandem mass spectrometric (LC–MS–MS) method was developed to determine pantoprazole sodium (PNT) in human urine. After solid-phase extraction with SPE cartridge, the urine sample was analysed on a C18 column (symmetry 3.5 μm; 75 mm × 4.6 mm i.d) interfaced with a triple quadrupole tandem mass spectrometer. Positive electrospray ionization was employed as the ionization source. The mobile phase consisted of acetonitrile–water (90:10, v/v). The method was linear over a concentration range of 1–100 ng mL−1. The lower limit of quantitation was 1 ng mL−1. The intra-day and inter-day relative standard deviation across three validation runs over the entire concentration range was <10.5%. The accuracy determined at three concentrations (8.0, 50.0 and 85.0 ng mL−1 PNT) was within ±1.25% in terms of relative errors.
3
Bhavesh H. Patel, Bhanubhai N. Suhagia, Madhabhai M. Patel et al. · Chromatographia · 2007 · 41 citations