Concepedia

Publication | Closed Access

Comparison of Cirrus OCT and Stratus OCT on the Ability to Detect Localized Retinal Nerve Fiber Layer Defects in Preperimetric Glaucoma

116

Citations

28

References

2010

Year

Abstract

There were no significant differences between the AUROCs for Cirrus and Stratus OCT, indicating that the two devices have similar diagnostic potentials in preperimetric glaucoma. After comparison with their normative databases, Cirrus OCT had generally higher sensitivities; however, this was largely at the cost of lower specificities than Stratus OCT.

References

YearCitations