Instrumentation Science & Technology · 1977 · 16 citations · 10 references
Natural SciencesBiochemistryChemistryBiophysical ChemistryMolecular BiophysicsSpectroscopyBiomolecular ThermodynamicsMolecular SpectroscopyMolecular ThermodynamicsOptical SpectroscopyRaman SpectroscopySpectroscopy Of BiomoleculesProtein Conformation ChangeRise TimeTemperature Jump KineticsTemperature-jump Techniques
ABSTRACT A Raman laser temperature-jump apparatus with highly sensitive (better than 10 −3 O.D. units) and very fast (rise time about 10 ns) pulsed spectrophotometric detection is described. A temperature jump of over 6°C is achieved within 20 ns in a special variable-path flow microcell. To illustrate the capabilities of the apparatus the kinetics of the conformation change of adenylyl 3‘→5’ adenosine in aqueous solution has been studied.
10
Technique of organic chemistry
Journal of Molecular Structure · 1970 · 983 citations
Martin Karplus, D. L. Weaver · Nature · 1976 · 628 citations
Dietmar Pörschke, Manfred Eigen · Journal of Molecular Biology · 1971 · 434 citations