Journal of the American Society for Mass Spectrometry · 2006 · 77 citations · 25 references
PharmacyChemistryPharmaceutical AnalysisAnalytical ChemistryMass Spectrometric AnalysisSpectroscopic AnalysisInstrumental AnalysisIonization TechniquesMass AnalyzersMass SpectrometryMass Spectrometry Data AnalysisDesorptionElectrospray IonizationFragmentationIon Source DesignIon Source PhysicsMass Spectrometric Drug AnalysisMass Spectrometric QuantificationRobustness
A MALDI ion source on a triple quadrupole mass spectrometer constructed for the purpose of obtaining high speed quantitative measurements on drugs and other low molecular weight compounds is described. Particular attention is given to the ion generation and transport phenomena that affect analysis speed, throughput, and practical instrument robustness. In this regard parameters that affect desorption speed, beam spreading, ion flight times, sensitivity, signal-to-noise, ion fragmentation, sample carry-over, and instrument contamination are examined and experimental results are provided. MALDI and electrospray sensitivity is compared, to provide a practical frame of reference.