IEEE Transactions on Nuclear Science · 2012 · 21 citations · 15 references
EngineeringNuclear EngineeringNuclear MedicineDiagnostic ImagingDiagnostic Nuclear MedicineNuclear InstrumentationDetector SystemsDetector TechnologyMedical ImagingMedical Imaging InstrumentationRadiation Imaging DetectorsScintillation DetectorsDetector DesignDetector PositioningDetectorsImaging DetectorsNuclear Medicine DetectorsPixel DetectorsScintillator Detector ResolutionSodium Iodide Detectors
NaI(Tl) pixels and decoded by an array of 51 mm diameter single-anode PMTs. Several configurations, utilizing two types of PMTs, were evaluated using a collimated beam source to measure positioning accuracy directly. Good pixel separation was observed, with correct pixel identification ranging from 60 to 72% averaged over the entire area of the modules, depending on the PMT type and configuration. This translates to a significant improvement in positioning accuracy compared to continuous slab detectors of the same thickness, along with effective reduction of "dead" space at the edges. The observed 10% average energy resolution compares well to continuous slab detectors. The combined performance demonstrates the suitability of pixelated detectors decoded with a relatively small number of medium-sized PMTs as a cost-effective approach for high resolution clinical SPECT applications, in particular those involving curved detector geometries.
15
Hal O. Anger · Review of Scientific Instruments · 1958 · 791 citations
Recent Advances in SPECT Imaging
Mark T. Madsen · Journal of Nuclear Medicine · 2007 · 333 citations · Full text
Fábio Esteves, Paolo Raggi, Russell D. Folks et al. · Journal of Nuclear Cardiology · 2009 · 263 citations · Full text