The Improvement of Joule Balance NIM-1 and the Design of New Joule Balance NIM-2

Zhengkun Li, Zhonghua Zhang, Qing He, Bing Han, Yunfeng Lu, Jinxin Xu, Shisong Li, Chen Li, Gang Wang, Tao Zeng,

IEEE Transactions on Instrumentation and Measurement · 2015 · 15 citations · 9 references

Concepts

Abstract

The development of the joule balance method to measure the Planck constant, in support of the redefinition of the kilogram, has been going on at the National Institute of Metrology of China (NIM) since 2007. The first prototype has been built to check the feasibility of the principle. In 2011, the relative uncertainty of the Planck constant measurement at NIM is 7.7 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-5</sup> . Self-heating and swing of the coils are the main uncertainty contributions. Since 2012, some improvements have been made to reduce these uncertainties. The relative uncertainty of the joule balance is reduced to 7.2 × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-6</sup> at present. The Planck constant measured with the joule balance is h = 6.6261041(470) × 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-34</sup> Js. The relative difference between the determined h and the CODATA2010 recommendation value is 5×10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-6</sup> . Further improvements are still being carried out on the NIM-1 apparatus. At the same time, the design and construction of a brand new and compact joule balance NIM-2 are also in progress and presented here.

References