Review of Scientific Instruments · 1997 · 15 citations · 0 references
Natural SciencesAstronomyPhysicsInstrumentationPlasma PhysicsSpectroscopyEmission SpectroscopyHigh-temperature PlasmasInfrared SpectroscopyOptical SpectroscopyUltraviolet InstrumentationUltraviolet SpectroscopyAbsorption Line BroadeningCharge-coupled Device DetectorsFar-ultraviolet SpectroscopyPlasma Emission SpectroscopySpectral Line BroadeningSpectral-line Surveys
We have constructed a simple, low-cost charge-coupled device (CCD) spectrometer capable of both high resolution (Δλ⩽0.015 nm) and large bandpass (110 nm with Δλ∼0.3 nm). These two modes of operation provide two broad areas of capability for plasma spectroscopy. The first major application is measurement of emission line broadening; the second is emission line surveys from the ultraviolet to the near infrared. Measurements have been made on a low-temperature plasma produced by a miniature electrostatic plasma source and the high-temperature plasma in the Madison Symmetric Torus reversed-field pinch. The spectrometer is a modified Jarrell–Ash 0.5 m Ebert–Fastie monochromator. Light is coupled into the entrance slit with a fused silica fiber optic bundle. The exposure time (2 ms minimum) is controlled by a fast electro-mechanical shutter. The exit plane detector is a compact and robust CCD detector developed for amateur astronomy by Santa Barbara Instrument Group. The CCD detector is controlled and read out by a Macintosh® computer. This spectrometer is sophisticated enough to serve well in a research laboratory, yet is simple and inexpensive enough to be affordable for instructional use.