Analytical Chemistry · 2011 · 55 citations · 43 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSpectroscopic AnalysisDetector SystemsInstrumental AnalysisIonization TechniquesLaser SpectroscopyMass Spectrometric Elemental AnalysisMass SpectrometryExplosivesExplosives AnalysisExplosives DetectionMass Spectral IdentificationMass Spectrometric IdentificationMeasurementSpectral SignaturesVapor Pressure
The mass spectral detection of low vapor pressure, inorganic-based explosive signatures including ammonium nitrate, chlorate, perchlorate, sugar, and the constituents contained within black powder are reported using laser electrospray mass spectrometry. The ambient pressure mass spectrometry technique combining nonresonant, femtosecond laser vaporization with electrospray postionization revealed primary and secondary signatures for trace quantities of inorganic explosives. A mixture of complexation agents in the electrospray solvent enabled the simultaneous detection of vaporized cations, anions, and neutrals in a single measurement. An offline classifier discriminated the inorganic-based explosives based on the mass spectral signatures resulting in high fidelity identification.
43
Ismael Cotte-Rodríguez, Zoltán Takáts, Nari Talaty et al. · Analytical Chemistry · 2005 · 345 citations
William Craig Byrdwell, W. E. Neff · Rapid Communications in Mass Spectrometry · 2002 · 203 citations