Review of Scientific Instruments · 2004 · 25 citations · 4 references
Natural SciencesNuclear EngineeringPhysicsInstrumentationPlasma PhysicsSpectroscopyAtomic SpectroscopyPlasma DiagnosticsCalibration TechniquesImpurity MeasurementsMeasurement AccuracyPolarizationPolarization CharacteristicsSignal-to-noise AnalysisSignal-to-noise AssessmentSignal-to-noise Ratio
An overview is given of the present design of the active beam spectroscopy diagnostics for the ITER. Present spatial resolution and signal-to-noise indicate that, in principle, all proposed measurements are possible covering the entire plasma minor radius. Calculations show that the mirror based periscopes and the impurity coating of the first mirror affect the signal strength and the polarization characteristics of the measured spectra having an impact in the measurement accuracy. On-line calibration techniques and methods to access the first mirror status are addressed.
4