IEEE Transactions on Nuclear Science · 1988 · 94 citations · 4 references
EngineeringBioengineeringBiomedical ImagingDiagnostic ImagingMicrofabricationMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsRadiation Imaging DetectorsX-ray DetectionX-ray DetectorsGainHigh-resolution X-ray DetectorsLinearityMicrochannel Plate DetectorsMicrochannel Plate ReadoutMicrochannel PlatesPosition-sensitive ReadoutResolutions
A two-dimensional delay line readout to be used with large-diameter (127-mm) microchannel plates (MCPs) has been developed. High spatial resolutions (25 mu m FWHM for an MCP gain of 2*10/sup 7/) and excellent linearity of position readout (less than +or-65- mu m deviation from linearity) have been achieved in a test setup using 30-mm-diameter MCPs coupled to 140-mm*140-mm delay lines. Single-pass delays of only 176 ns allow system deadtimes of less than 1 mu s. This type of readout will be applied to imaging X-ray detectors to be used in protein crystallography.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
4
A position sensitive microchannelplate detector using a delay line readout anode
H. U. Keller, G. Klingelhöfer, E. Kankeleit · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 1987 · 47 citations
Engineering, Electrical Engineering, Microelectromechanical Systems +13