Magnetic Resonance in Medicine · 1999 · 65 citations · 13 references
Natural SciencesBiomedical EngineeringBiomedical ImagingDiagnostic ImagingThermal AnalysisMagnetic CharacterizationMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationSuperconducting Magnet SystemsSuperconducting MaterialsCryogenic EngineeringField HomogeneityHigh-temperature Superconducting Magnet TechnologyMagnet Cooling SystemsMagnetic Field HomogeneityMagnetic ResonanceMagnetic Resonance Probe DesignThermal Stability
The design and operation of a high-temperature superconducting (HTS) probe for magnetic resonance microscopy (MRM) at 400 MHz are presented. The design of the probe includes a Helmholtz coil configuration and a stable open-cycle cooling mechanism. Characterization of coil operating parameters is presented to demonstrate the suitability of cryo-cooled coils for MRM. Specifically, the performance of the probe is evaluated by comparison of signal-to-noise (SNR) performance with that of a copper Helmholtz pair, analysis of B1 field homogeneity, and quantification of thermal stability. Images are presented to demonstrate the SNR advantage of the probe for typical MRM applications.
13
A High-Temperature Superconducting Receiver for Nuclear Magnetic Resonance Microscopy
Robert D. Black, T. A. Early, P. B. Roemer et al. · Science · 1993 · 205 citations