Science · 1993 · 205 citations · 18 references
EngineeringBiomedical EngineeringBiophysicsBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance SpectroscopyCoil DesignMagnetic Resonance PhysicsNuclear Magnetic Resonance SpectroscopyNuclear Magnetic Resonance StudiesReceiver NoiseSignal-to-noise RatioSignal-to-noise Ratio AnalysisSuperconducting Receivers
A high-temperature superconducting-receiver system for use in nuclear magnetic resonance (NMR) microscopy is described. The scaling behavior of sources of sample and receiver-coil noise is analyzed, and it is demonstrated that Johnson, or thermal, noise in the receiver coil is the factor that limits resolution. The behavior of superconductors in the environment of an NMR experiment is examined, and a prototypical system for imaging biological specimens is discussed. Preliminary spin-echo images are shown, and the ultimate limits of the signal-to-noise ratio of the probe are investigated.
18
Thermal Agitation of Electricity in Conductors
J. B. Johnson · Physical Review · 1928 · 1.8K citations
F. Bloch, W. W. Hansen, M. E. Packard · Physical Review · 1946 · 848 citations · Full text
The intrinsic signal‐to‐noise ratio in NMR imaging
W. A. Edelstein, Gary H. Glover, C.J. Hardy et al. · Magnetic Resonance in Medicine · 1986 · 754 citations