Natural SciencesChemistryElectromagneticsElectromagnetismSpectroscopyDielectric PropertiesDielectric ResponseDielectricsElectron Paramagnetic ResonanceElectron Spin Resonance SpectroscopyMagnetic Resonance SpectroscopyAqueous Cell DesignDielectric CharacterizationDielectric Loss MechanismsDielectric SpectroscopyHigh-frequency Electron Paramagnetic ResonanceMagnetic FluxMagnetic Flux DensityProbe Design
If an aqueous sample cell is arranged such that the electric field vectors in the cavity are substantially perpendicular to the surface of the cell, a good cavity Q is observed and high quality EPR signals can be obtained. Polarization charges induced on the surface of the sample cause a very marked reduction of the electric field strength in the sample compared with that in the cavity, and the dielectric loss is small. The rf magnetic flux density B as determined by EPR experiments is substantially unchanged, indicating that ▿×E in the sample is unchanged.