Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 1994 · 16 citations · 0 references
Natural SciencesAstronomyOpticsRadiation DetectionRadiation MeasurementRadiographic ImagingX-ray DetectionX-ray DetectorsX-ray OpticsX-ray PolarimetryDetector DesignImaging X-ray PolarimetryPolarimeter DesignPolarization-sensitive DetectorsX-ray Imaging PhysicsX-ray Polarization Measurements
An optical imaging chamber can be used to detect x-rays by imaging the tracks of photoelectrons ejected when the x-rays are absorbed in the detector volume. These tracks contain information about the location of the x-ray interaction point and its polarization. In the lab, we have obtained a modulation factor of 30% for 60 keV polarized x-rays. Here we discuss preliminary work done towards building a large area hard x-ray imaging polarimeter which will be able to measure x-ray polarizations from bright cosmic x-ray sources at energies between 40 keV and 100 keV.