Review of Scientific Instruments · 1990 · 17 citations · 10 references
Natural SciencesNuclear EngineeringPhysicsFusion EngineeringNuclear InstrumentationPlasma PhysicsHigh-temperature PlasmasPlasma ConfinementPlasma DiagnosticsTokamak EngineeringFusion Reactor EngineeringFusion Reactor Physics
Unmonitored changes in the transmission of viewing windows owing to deposited films can produce errors in Thomson scattering temperature measurements. This effect is illustrated by a recent run on the Tokamak Fusion Test Reactor (TFTR) where apparent errors of over 25% in peak temperatures of 9 keV owing to carbon films were noted. Since coatings can also be removed by hydrogen discharges, the transmission of a window may change with time, resulting in variable errors in the temperature. It is proposed that these changes be monitored by calibration in situ with the aid of a low-pressure hydrogen glow discharge.
10
Laser light scattering in laboratory plasmas
D. Evans, J. Katzenstein · Reports on Progress in Physics · 1969 · 418 citations
Natural Sciences, Physics, Optics +16
The molecular spectrum of hydrogen and its isotopes
G. H. Dieke · Journal of Molecular Spectroscopy · 1958 · 254 citations
Natural Sciences, Chemistry, Physics +12
High-temperature plasmas in a tokamak fusion test reactor
J. Strachan, M. Bitter, A. T. Ramsey et al. · Physical Review Letters · 1987 · 250 citations · Full text
TFTR Thomson scattering system
D. Johnson, D. Dimock, B. Grek et al. · Review of Scientific Instruments · 1985 · 54 citations
Natural Sciences, Physics, Optics +12