Effect of changes in viewing window transmission on high-temperature Thomson scattering data

D. H. Mcneill

Review of Scientific Instruments · 1990 · 17 citations · 10 references

DOIFull text

Open access

Concepts

Abstract

Unmonitored changes in the transmission of viewing windows owing to deposited films can produce errors in Thomson scattering temperature measurements. This effect is illustrated by a recent run on the Tokamak Fusion Test Reactor (TFTR) where apparent errors of over 25% in peak temperatures of 9 keV owing to carbon films were noted. Since coatings can also be removed by hydrogen discharges, the transmission of a window may change with time, resulting in variable errors in the temperature. It is proposed that these changes be monitored by calibration in situ with the aid of a low-pressure hydrogen glow discharge.

References

10