Evaluation of a dual-arm Archimedean spiral array for microwave hyperthermia

Jessi E. Johnson, D. Neuman, Paolo F. Maccarini, Titania Juang, Paul R. Stauffer, Peter Turner

International Journal of Hyperthermia · 2006 · 48 citations · 33 references

DOIFull text

Open access

Concepts

Abstract

PURPOSE: This effort describes a third-party performance evaluation of a novel, commercial, dual-armed Archimedean spiral array hyperthermia applicator. The applicator is analysed for its ability to couple efficiently into muscle equivalent phantom loads, operate over a broad bandwidth to help accommodate variable tissue properties and generate predictable and repeatable SAR contours that are adaptable to clinically probable disease shapes. MATERIALS AND METHODS: Characterization of the applicator includes E-field and return-loss measurements in liquid muscle tissue-equivalent phantom, as well as comparison of 'treatment-planning' simulations of several possible array SAR patterns with measured SAR from non-coherently driven spiral array antennae. RESULTS: The applicator demonstrates a reasonably low return loss over a large bandwidth and the ability to generate a very uniform heating pattern. Ability to adjust SAR contours spatially to fit specific shapes is also demonstrated. CONCLUSIONS: This device should prove a welcome addition to a currently limited set of superficial heating applicators to provide controllable heating of superficial tissue disease.

References