IEEE Transactions on Biomedical Engineering · 1992 · 79 citations · 8 references
EngineeringBiomedical EngineeringOncologyAntennasMedical PhysicsTreatment PlanningMedical InstrumentationMedical Radiation ShieldingMicrostrip AntennasBlood FlowDesign Tradeoff AnalysisEngineering Design ProcessMicrostrip Array AntennasPower DepositionPower Deposition ProfilesSolid PhantomsTissue-equivalent Materials
The optimal treatment with hyperthermia of superficially located tumors which involve large surface areas requires applicators which can physically conform to body contours, and locally alter their power deposition patterns to adjust for nonuniform temperature caused by tissue inhomogeneities and blood flow variations. A series of 915 MHz microstrip array applicators satisfying these criteria have been developed and clinically tested. Clinical and engineering design tradeoffs for practical devices are discussed. Measurements taken in tissue equivalent phantoms and a summary of our clinical experiences with these microstrip arrays are presented.
8
A Multi-Element Microstrip Antenna for Local Hyperthermia
Eiji Tanabe, A. McEuen, C.S. Norris et al. · 2005 · 25 citations