Biomedical Optics Express · 2015 · 91 citations · 38 references
EngineeringSurface-enhanced Raman ScatteringBiomedical EngineeringSeparation-free MeasurementsRenal FunctionBiosensing SystemsBioanalysisAnalytical ChemistryUrine SamplesBioimagingClinical ChemistryNanosensorMouse ModelMolecular ImagingBiophysicsChromatographyPlasmonic MaterialBiochemistryBiomedical AnalysisUrologyCreatinine Concentration MeasurementsBiomedical DiagnosticsSpectroscopyUrine Creatinine ConcentrationMedicineSpectroscopic Method
We report a novel reagent- and separation-free method for urine creatinine concentration measurement using stamping surface enhanced Raman scattering (S-SERS) technique with nanoporous gold disk (NPGD) plasmonic substrates, a label-free, multiplexed molecular sensing and imaging technique recently developed by us. The performance of this new technology is evaluated by the detection and quantification of creatinine spiked in three different liquids: creatinine in water, mixture of creatinine and urea in water, and creatinine in artificial urine within physiologically relevant concentration ranges. Moreover, the potential application of our method is demonstrated by creatinine concentration measurements in urine samples collected from a mouse model of nephritis. The limit of detection of creatinine was 13.2 nM (0.15 µg/dl) and 0.68 mg/dl in water and urine, respectively. Our method would provide an alternative tool for rapid, cost-effective, and reliable urine analysis for non-invasive diagnosis and monitoring of renal function.
38
Concluding Remarks : Surface enhanced Raman scattering
Michael J. Natan · Faraday Discussions · 2006 · 396 citations
Urine analysis by laser Raman spectroscopy
W. Ranjith Premasiri, Richard H. Clarke, M. Womble · Lasers in Surgery and Medicine · 2001 · 127 citations