Physiological Measurement · 2003 · 41 citations · 19 references
Veterinary MedicineOrthopaedic SurgerySurgeryOrthopedicsUltrasonographyUltrasoundCompartment SyndromeDiagnostic UltrasonographyMusculoskeletal UltrasonographyUltrasonicsUltrasound InstrumentationCompartment Syndrome AssessmentCompartment Syndrome DetectionLeg Compartment SyndromeMuscle Compartment AnatomyNoninvasive Compartment MonitoringNoninvasive DiagnosticsNoninvasive Imaging DiagnosisSub-micrometer Metrology
An ultrasonic device for the diagnosis of acute compartment syndrome is described and results on six human cadaveric legs are presented. The ultrasonic device uses a pulsed phase locked loop (PPLL) to measure sub-micrometer displacements of the fascia wall. These displacements occur as a result of volume expansion of the muscle compartment of the lower leg and are related to changes in intramuscular pressure (IMP). In the cadaveric tests, the PPLL detected changes in compartment diameter resulting from IMP changes of 1 mmHg and from infusions of 0.25 ml saline increments. Based on these results, the ultrasonic PPLL appears to have the potential to become a low-cost, portable and noninvasive alternative to current methods for diagnosing acute compartment syndrome.
19
Michael Heckman, Thomas E. Whitesides, Scott R. Grewe et al. · Journal of Bone and Joint Surgery · 1994 · 218 citations
Akio Kobayashi, M. A. R. Freeman, W. Bonfield et al. · Journal of Bone and Joint Surgery - British Volume · 1997 · 196 citations