Review of Scientific Instruments · 2008 · 27 citations · 5 references
Natural SciencesPhysicsMicroelectronicsPlasma PhysicsSpectroscopyHigh-temperature PlasmasPlasma DiagnosticsPlasma SourcesElectron Cyclotron Resonance Plasma SourcesIon Source PhysicsMicrowave PlasmasPlasma BreakdownPlasma ProcessesPulsed OperationPulsed Plasma Sources
A careful study of pulsed mode operation of the PHOENIX electron cyclotron resonance (ECR) ion source has clearly demonstrated the reality of an unexpected transient current peak, occurring at the very beginning of the plasma breakdown. This regime was named the preglow, as an explicit reference to the afterglow occurring at the microwave pulse end. After the transient preglow peak, the plasma regime relaxes to the classical steady state one. Argon preglow experiments performed at LPSC are presented. A theoretical model of ECR gas breakdown in a magnetic trap, developed at IAP, showing satisfactory agreement with the experimental results is suggested.
5
ECRIS development for stable and radioactive pulsed beams