Testing by Sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory: a comparison

R. Evans, M.M. Davidson, Lyann Sim, A. Hay

Journal of Clinical Pathology · 2006 · 34 citations · 6 references

DOIFull text

Open access

Abstract

A large proportion of the samples tested in routine diagnostic microbiology laboratory are urine samples. The gold standard is bacterial culture, but a high proportion of samples cultured are negative. Unnecessary testing can be reduced and an improved service provided by an effective screening test. The Sysmex UF-100 flow cytometer has been developed to count cells and casts accurately in urine samples. Its performance in a screening test was compared with bacterial culture by using 1005 consecutive urine samples, and cut-off criteria were established. Cut-off values of 3000 bacteria/microl and 111 WBC/microl provided the best discrimination. Of 1005 samples, 606 (60%) would be cultured. Sixteen samples that were not selected according to these criteria were culture positive. This was considered acceptable for our routine use. The use of a testing algorithm incorporating the Sysmex UF-100 flow cytometer has improved the quality and efficiency of urine testing within the routine microbiology laboratory.