Journal of Clinical Pathology · 2006 · 34 citations · 6 references
A large proportion of the samples tested in routine diagnostic microbiology laboratory are urine samples. The gold standard is bacterial culture, but a high proportion of samples cultured are negative. Unnecessary testing can be reduced and an improved service provided by an effective screening test. The Sysmex UF-100 flow cytometer has been developed to count cells and casts accurately in urine samples. Its performance in a screening test was compared with bacterial culture by using 1005 consecutive urine samples, and cut-off criteria were established. Cut-off values of 3000 bacteria/microl and 111 WBC/microl provided the best discrimination. Of 1005 samples, 606 (60%) would be cultured. Sixteen samples that were not selected according to these criteria were culture positive. This was considered acceptable for our routine use. The use of a testing algorithm incorporating the Sysmex UF-100 flow cytometer has improved the quality and efficiency of urine testing within the routine microbiology laboratory.
6
Childhood reflux and urinary infection: a follow-up of 10-41 years in 226 adults
J. M. Smellie, N Prescod, Pamela J. Shaw et al. · Pediatric Nephrology · 1998 · 314 citations
Pathogenesis of urinary tract infections: an update
T. M. Hooton · Journal of Antimicrobial Chemotherapy · 2000 · 213 citations · Full text
F. Manoni, S. Valverde, F Antico et al. · Clinical Microbiology and Infection · 2002 · 42 citations · Full text
Urogenital Radiology, Second-generation Cytometer Uf-100, Urology +9