Review of Scientific Instruments · 1997 · 69 citations · 26 references
High ResolutionEngineeringPfi-pe StudiesSynchrotron Radiation ResearchElectron SpectroscopyIon BeamInstrumentationIon EmissionUndulator Synchrotron RadiationPhysicsAtomic PhysicsHigh Resolution ThresholdPhotoelectric MeasurementSynchrotron RadiationMulti-bunch Synchrotron RadiationNatural SciencesSpectroscopyApplied PhysicsOptoelectronics
We have demonstrated a resolution of 0.8 meV [full width at half-maximum (FWHM)] for threshold photoelectron measurements using a steradiancy-type zero kinetic energy photoelectron (ZEKE-PE) analyzer and the high resolution monochromatized vacuum ultraviolet (VUV) undulator synchrotron radiation of the chemical dynamics beamline at the advanced light source (ALS). Using this high resolution ZEKE-PE energy analyzer to filter prompt electrons and by employing a proper voltage pulsing scheme adapted to the timing structure of the ALS, we have achieved a resolution of 0.5 meV (FWHM) for pulsed field ionization photoelectron (PFI-PE) measurements with little contamination from prompt photoelectrons produced from direct photoionization and autoionizing processes. The experiment scheme presented here is generally applicable to PFI-PE studies using multi-bunch VUV synchrotron radiation at other synchrotron radiation facilities.
26
<i>Molecular Photoelectron Spectroscopy</i>
D. W. Turner, Christopher A. Baker, A. D. Baker et al. · Physics Today · 1972 · 1.2K citations
The ionization energy of nitric oxide
Georg Reiser, Wieland Habenicht, Klaus Müller‐Dethlefs et al. · Chemical Physics Letters · 1988 · 342 citations
Klaus Müller‐Dethlefs, Michael Sander, Edward W. Schlag · Zeitschrift für Naturforschung A · 1984 · 242 citations · Full text