Physics in Medicine and Biology · 1978 · 54 citations · 13 references
EngineeringNuclear EngineeringNuclear MedicineElemental AnalysisNuclear InstrumentationRadiation MeasurementActivation AnalysisActivity MeasurementDetector SystemsDosimetry InstrumentationNeutron Activation AnalysisPhantom DosimetryNeutron ActivationNeutron Detector SystemsRadiochemical Neutron Activation AnalysisSolid Phantoms
A system has been established for determining the total body contents of calcium, phosphorus, sodium, chlorine and nitrogen by neutron activation analysis using sealed-tube neutron generators and a shadow-shield whole-body counter. The system was calibrated using three anthropomorphic phantoms of different sizes, filled with various mixtures of the activatable elements in amounts similar to those in man. The reproducibility determined from replicate measurements, was on average +/- 2.9% (SD). The average uncertainty in absolute measurements was estimated to be +/- 4.2% (68% confidence interval).
13