IEEE Transactions on Very Large Scale Integration (VLSI) Systems · 2015 · 16 citations · 15 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBiomedical Signal ProcessingLow-power DesignAnalog Integrated CircuitsBiomedical InstrumentationCmos Circuit DesignDigital Integrated CircuitsElectrocardiogram AnalysisElectrocardiographyAnalog Circuit DesignAnalog-to-digital Converter DesignCmos Analog Circuit DesignCmos Analog Integrated CircuitsDigital FeedbackFront-end ArchitectureLow-noise DesignPower Consumption
This paper presents a new power-efficient electrocardiogram acquisition system that uses a fully digital architecture to reduce the power consumption and chip area. The proposed architecture is compatible with digital CMOS technology and is capable of operating with a low supply voltage of 0.5 V. In this architecture, no analog block, e.g., low-noise amplifier (LNA), and filters, and no passive elements, such as ac coupling capacitors, are used. A moving average voltage-to-time converter is used, which behaves instead of the LNA and antialiasing filter. A digital feedback loop is employed to cancel the impact of the dc offset on the circuit, which eliminates the need for coupling capacitors. The circuit is implemented in 0.18-um CMOS process. The simulation results show that the front-end circuit consumes 274 nW of power.
15
Qinwen Fan, Fabio Sebastiano, Johan H. Huijsing et al. · IEEE Journal of Solid-State Circuits · 2011 · 446 citations
Engineering, Biomedical Engineering, Electrical Engineering +17