Comparison of 4 Airway Devices on Cervical Spine Alignment in a Cadaver Model With Global Ligamentous Instability at C5–C6

38

Citations

26

References

2011

Year

Abstract

In a cadaver model of C5-C6 instability, the greatest amount of motion was caused by the most commonly used intubation device, the Macintosh blade. Intubation with the Lightwand resulted in significantly less motion in all tested parameters (other than ML translation) as compared with the Macintosh blade. It should also be noted that the Airtraq caused less motion than the Macintoshblade in 3 of the 6 tested planes. There were no significant differences in failure rate or the amount of time it took to successfully intubate in comparing these techniques. We therefore recommend the use of the Lightwand, followed by the Airtraq, in the setting of a presumed unstable cervical spine injury over the Macintosh laryngoscope.