PubMed · 1983 · 458 citations · 0 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryJoint Replacement SurgeryOrthopedic BiomechanicsOrthopedicsArthroplasty BiomechanicsKnee ArthroplastyKnee BiomechanicsLower Limb AlignmentEccentric LoadingEccentric Loading EffectsKnee Joint AlignmentKnee Valgus AlignmentKnee Valgus DeformityKnee Varus AlignmentTotal Knee Arthroplasty BiomechanicsVarus Loading
The initial experience with the Freeman-Swanson (ICLH) total knee arthroplasty (1971-1975) led to changes of tibial component size, alignment, instruments, and surgical technique in 1975. Certain patients in this initial series, however, have shown excellent long-term (5 1/2-9 years) results. In patients with overall alignment between 1 degrees and 13 degrees valgus, the success rate was 89%. In patients with knees aligned between neutral and 8 degrees varus, the success rate was only 14%. This startling difference in long-term results was investigated by using eccentric loading tests to simulate varus and valgus malalignment. Knee arthroplasties failed in eccentric loading at very low loads. Failure by lift-off of the unloaded side occurred at 9%-18% and failure of bone due to compression on the loaded side occurred at 34%-51% of the load at failure under ideal uniform loading conditions. Thus, both the long-term clinical study and the laboratory analysis demonstrate the importance of proper overall alignment in total knee arthroplasty.