Review of Scientific Instruments · 1999 · 79 citations · 10 references
Natural SciencesChemistryPhysicsAnalytical ChemistryPolymer ChemistrySpectroscopyElectron Paramagnetic ResonanceElectron Spin Resonance SpectroscopyMagnetic Resonance SpectroscopyMicrowave MeasurementsMicrowave SpectroscopyOptical SpectroscopyBandwidthContinuous-wave Electron Paramagnetic ResonanceContinuous-wave Electron Spin ResonanceHeterodyne DetectionHigh-field Electron Spin ResonanceHigh-frequency Electron Paramagnetic ResonanceHigh-frequency Electron Spin ResonanceSpectrometer Design
We present design details of and first measurements with a novel continuous wave (cw) high-field/high-frequency electron paramagnetic resonance spectrometer operating at a microwave frequency of 360 GHz and a magnetic field of up to 14 T. The spectrometer design incorporates a heterodyne mixer detection scheme with a quasi-optical transmission line and a bimodal induction mode Fabry–Perot cavity. First cw experiments on polycrystalline 1,1-diphenyl-2-picryl-hydrazyl and bisdiphenylene-β-phenylallyl benzolate in polystyrene at room temperature and 4-hydroxy-2,2,6,6-tetramethylpiperidine-1-oxyl in frozen solution at 190 K demonstrate the high Zeeman resolution achievable and allow an estimate of the present detection sensitivity of 4×109 spins/G at a detection bandwidth of 1 Hz.
10
Electron spin-echo spectroscopy at 95 GHz
R. Weber, J.A.J.M. Disselhorst, L.J Prevo et al. · Journal of Magnetic Resonance (1969) · 1989 · 69 citations