Journal of Liquid Chromatography · 1983 · 24 citations · 8 references
Natural SciencesBiochemistryAnalytical ChemistryChromatographic AnalysisClinical BiochemistryBioanalytical Method ValidationChromatographyClinical AnalysisVitamin AnalysisVitamin BiochemistryVitamin ChemistryElectrochemical AssaysPlasma Ascorbate MeasurementPlasma Ascorbic Acid LevelsReversed-phase Columns
Abstract Following the stabilization of the plasma samples with HClO4 and EDTA, the samples could be directly analyzed by HPLC using electrochemical detection and reversed-phase columns. The accuracy and precision of the method was evaluated using plasma samples spiked with ascorbic acid (10 μg/ml) and the results were also compared to the classical colorimetric procedure. Dehydroascorbic (5 μg/ml) was determined in plasma samples using UV detection following derivatization at room temperature for 45 minutes with o-phenylenediamine.
8
Histamine and Ascorbic Acid in Human Blood
C. Alan B. Clemetson · Journal of Nutrition · 1980 · 77 citations