Journal of Biomechanical Engineering · 1990 · 94 citations · 0 references
MedicineNephrologyPhysiologyDialysisHemodialysisTransport PhenomenaDialyzer MembranesHemodynamicsMembrane FiltrationRenal ClearanceRenal Drug ClearanceSolute ClearanceBlood FlowCoupled Mass TransferDialysis Mass TransferDiffusive TransportDrug Clearance By HemodiafiltrationMass Transfer MechanismsSolute Mass TransferUltrafiltration
The mass transfer in a hemodialyser in the presence of combined dialysis and ultrafiltration has been calculated by integration of mass fluxes across the boundary layers in blood and dialysate phase taking into account the partial rejection of solute as well as changes in local blood flow and ultrafiltration flux along the membrane. Clearances of creatinin, vitamin B12, and myoglobin have been calculated as a function of blood and ultrafiltrate flow rate and were found to be in good agreement with in vitro measurements. The data suggest the following empirical correlation for the hemodiafiltration clearance.