Cytometry Part B Clinical Cytometry · 2009 · 12 citations · 20 references
We have designed and evaluated the performance of a simple, rapid, and affordable method for counting CD4(+) T-cells with the use of plastic microchips. This new system is an adaptation of a "no-lyse, no-wash," volumetric single platform assay, and absolute CD4(+) counts are determined with the use of a microscopic scanning cell counter. To assess the CD4(+) count test precision and linearity of the system, measured CD4(+) counts were compared with two other reference assays (single and dual platform flow cytometry) with the use of 123 clinical samples including samples obtained from 35 HIV-infected patients, and artificially diluted samples. A correlation between the results from the use of the new method and from the use of the two other reference assays was r = 0.98 for the clinical samples. A dilution test of the new method demonstrated a linearity of r >or= 0.99, with coefficients of variation <or=7.6% for all concentration levels. Our findings suggest that the new CD4(+) counting device can be potentially be applied for other diagnostic procedures that measure quantities of characteristic antigens or other materials on cells.
20
A H BECKETT, M ROWLAND, P TURNER · The Lancet · 1965 · 16.9K citations
AIDS epidemic update: December 1999.
PubMed · 2000 · 447 citations
Epidemiological Dynamic, Disease Outbreak, Aids Epidemic Update +22