Radiometric calibration of the in-flight blackbody calibration system of the GLORIA interferometer

C. Monte, B. Gutschwager, Albert Adibekyan, M. Kehrt, A. Ebersoldt, F. Olschewski, J. Hollandt

Atmospheric measurement techniques · 2014 · 16 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

Abstract. GLORIA (Gimballed Limb Observer for Radiance Imaging of the Atmosphere) is an airborne, imaging, infrared Fourier transform spectrometer that applies the limb-imaging technique to perform trace gas and temperature measurements in the Earth's atmosphere with three-dimensional resolution. To ensure the traceability of these measurements to the International Temperature Scale and thereby to an absolute radiance scale, GLORIA carries an on-board calibration system. Basically, it consists of two identical large-area and high-emissivity infrared radiators, which can be continuously and independently operated at two adjustable temperatures in a range from −50 °C to 0 °C during flight. Here we describe the radiometric and thermometric characterization and calibration of the in-flight calibration system at the Reduced Background Calibration Facility of the Physikalisch-Technische Bundesanstalt. This was performed with a standard uncertainty of less than 110 mK. Extensive investigations of the system concerning its absolute radiation temperature and spectral radiance, its temperature homogeneity and its short- and long-term stability are discussed. The traceability chain of these measurements is presented.

References