Minimizing Matrix Effects While Preserving Throughput in LC–MS/MS Bioanalysis

Zhengqi Ye, Hong Tsao, Hong Gao, Christopher L. Brummel

Bioanalysis · 2011 · 32 citations · 16 references

Concepts

Abstract

BACKGROUND: Phospholipids are known to cause matrix effects in LC-MS analysis and are not effectively removed by one of the most common method of sample preparation: organic solvent protein precipitation. The objective of this research is to minimize phospholipid interferences chromatographically. RESULTS: In this article we examine several chromatographic approaches and highlight the method we developed that allows for the rapid gradient separation of model drug molecules from phospholipids. CONCLUSION: The new approach (which utilizes a mixture of methanol and acetonitrile as the organic mobile phase on a 2.1 × 20 mm C18 column) minimized phospholipids-related matrix effects in the analysis of plasma samples prepared by protein precipitation and is suitable for high-throughput bioanalysis in drug discovery.

References

16