Optics Letters · 2010 · 269 citations · 11 references
Tissue ImagingEngineeringVascular ImageMedical ImagingOphthalmologyBiomedical OpticBiomedical ImagingVascular ImagingDoppler OctSpeckle VarianceBiomedical EngineeringBiophotonicsOptical Coherence TomographyMedical Image ComputingMedicineOptical ImagingSvoct Imaging
We optimize speckle variance optical coherence tomography (svOCT) imaging of microvasculature in high and low bulk tissue motion scenarios. To achieve a significant level of image contrast, frame rates must be optimized such that tissue displacement between frames is less than the beam radius. We demonstrate that higher accuracy estimates of speckle variance can enhance the detection of capillaries. These findings are illustrated in vivo by imaging the dorsal window chamber model (low bulk motion). We also show svOCT imaging of the nonstabilized finger (high bulk motion), using optimized imaging parameters, demonstrating better vessel detection than Doppler OCT.
11
Three dimensional optical angiography
Ruikang K. Wang, Steven L. Jacques, Zhenhe Ma et al. · Optics Express · 2007 · 625 citations · Full text
Real-time fiber-based multi-functional spectral-domain optical coherence tomography at 13 µm
B. Hyle Park, Mark C. Pierce, Barry Cense et al. · Optics Express · 2005 · 349 citations · Full text
Engineering, Coherence, Fiber Optics +17
Phase-resolved optical frequency domain imaging
Benjamin J. Vakoc, Seok Hyun Yun, Johannes F. de Boer et al. · Optics Express · 2005 · 297 citations · Full text