A new miniature x‐ray source for interstitial radiosurgery: Device description

Michael Dinsmore, K. J. Harte, Alan Sliski, D. O. Smith, P. M. Nomikos, M. J. Dalterio, A. J. Boom, W. F. Leonard, P. E. Oettinger, J. C. Yanch

Medical Physics · 1996 · 109 citations · 0 references

Concepts

Abstract

A device that generates low-energy x rays at the tip of a needle-like probe was developed for stereotactic interstitial radiosurgery. Electrons from a small thermionic gun are accelerated to a final energy of up to 40 keV and directed along a 3 mm outside diameter drift tube to a thin Au target, where the beam size is approximately 0.3 mm. All high-voltage electronics are in the probe housing, connected by low-voltage cable to a battery-operated control box. X-ray output, which is nearly isotropic, consists of a bremsstrahlung spectrum and several lines between 7 and 14 keV, with characteristic radiation contributing 15% of the total energy output. To date, 14 patients with metastatic brain tumors have been treated with this device.