Powder Diffraction · 2011 · 105 citations · 14 references
Natural SciencesMaterials EngineeringMaterials ScienceMaterials CharacterizationMaterials ChemistryMetrologyCrystallographyNeutron DiffractionPowder ProcessingX-ray CrystallographyX-ray DiffractionCrystallite Size DeterminationCrystallographic DefectsDiffraction AnalysisPowder CharacterizationPowder DiffractionPowder Diffraction AnalysisPowder Neutron DiffractionPrecursor Materials
This report describes SRM 660b, the third generation of this powder diffraction SRM used primarily for determination of the instrument profile function (IPF). It is certified with respect to unit-cell parameter. It consists of approximately 6 g LaB 6 powder prepared using a 11 B isotopically enriched precursor material so as to render the SRM applicable to the neutron diffraction community. The microstructure of the LaB 6 powder was engineered to produce a crystallite size above that where size broadening is typically observed and to minimize the crystallographic defects that lead to strain broadening. A NIST -built diffractometer, incorporating many advanced design features, was used to certify the unit-cell parameter of the LaB 6 powder. Both type A, statistical, and type B, systematic, errors have been assigned to yield a certified value for the unit-cell parameter of a =0.415691(8) nm at 22.5°C.
14
<i>X-Ray Diffraction in Crystals, Imperfect Crystals, and Amorphous Bodies</i>
A. Guinier, Paul Lorrain, Dorothée Sainte-Marie Lorrain et al. · Physics Today · 1964 · 1.4K citations
Natural Sciences, Chemistry, Physics +14