A high pressure fiber-optic reactor with charge-coupled device array ultraviolet-visible spectrometer for monitoring chemical processes in supercritical fluids

Fred Hunt, Hiroyuki Ohde, Chien M. Wai

Review of Scientific Instruments · 1999 · 24 citations · 8 references

Concepts

Abstract

A versatile spectroscopic reaction monitoring system for high-pressure liquids and supercritical fluids is described. The cell is designed to be attached to a charge-coupled device array ultraviolet-visible spectrometer via fiber optics, and has been shown to be capable of monitoring high speed reactions and rate of dissolution of a variety of compounds. The design is simple, inexpensive, rugged, and self-aligning. The cell has been tested up to 300 atm and 100 °C without failure.

References

8