Magnetic Resonance in Medicine · 2005 · 13 citations · 13 references
Natural SciencesRadiologyDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance BiophysicsImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationMedical ImagingBalanced Steady-state Free PrecessionEquilibriumRelaxation TimesResonance ScanningSpin PrecessionSpin-lattice RelaxationStarter Sequence DesignSteady-state Signal Formation
The dynamic equilibrium exploited by balanced steady-state free precession imaging develops slowly because its formation is dependent on both spin-spin and spin-lattice relaxation times. Attempting to image before steady state is established results in artifacts due to transient signal oscillations. Using a starter sequence to precondition the spin system can significantly reduce the delay before imaging. An improved design for a steady-state starter sequence is presented. The new sequence has the advantage of uniformly exciting the steady-state response for all resonance offsets and can be phase cycled to suppress banding artifacts.