Journal of Physics E Scientific Instruments · 1970 · 21 citations · 4 references
EngineeringNuclear EngineeringNuclear InstrumentationRadiation DetectionRadiation PhysicsDetector SystemsHighly Charged IonsChannel Electron MultipliersEnergy DependenceIon Detection EfficiencyRelative Efficiency
A determination has been made of the efficiency with which certain commercially available channel electron multipliers detect H+ and Ar+ ions. The three multipliers tested were Bendix types 4013 and 4039, which have 3 mm and 10 mm diameter funnels respectively, and a Mullard type B419 BL which has a 10 mm diameter funnel. The efficiency shows saturation for ions with energy above a few kilovolts and is relatively constant across the whole surface of the 10 mm funnels. The particular 3 mm funnel tested shows an energy dependent sensitivity across its surface.
4
Channeltron Fatigue and Efficiency for Protons and Electrons
A. Egidi, R. Marconero, G. Pizzella et al. · Review of Scientific Instruments · 1969 · 68 citations