Electrophoresis · 1999 · 40 citations · 31 references
PharmacyBiochemistryBioinstrumentationBiopharmaceuticsExcretionBioanalytical Method ValidationMetabolic EvaluationUrinary ExcretionAcetate BuffersCompetitive AthletesCreatinine Analysis In Body FluidsCreatinine-based BiomarkersMethod Detection LimitRapid Urine DiagnosticsReference Value ValidationSample Injection
Using capillary zone electrophoresis, the urine creatinine (uCr) assay was validated in extemporaneous diluted urine, both in healthy subjects and athletes, with the uCr concentration as a reference value to compare excretion rates of other metabolites in the same samples. The electrokinetic sample injection was carried out at 10 kV per 10 s; UV absorbance detection was at 254 nm. Using standard samples, the creatinine migration mean time in 100 mmol/L acetate buffer, pH 4.4, was 3.3+/-0.2 min; the repeatability for absolute migration mean time was 0.6% and peak height repeatability was 2.9%. The correlation coefficient of the standard curve was r = 0.999 and the detection limit was 23.1 micromol/L. Intra- and interassay coefficients of variation (CV) were 3.0 and 3.6%, respectively; recovery was 99+/-3% and linearity was r= 0.98. Normal urine samples were diluted 1:80 in run buffer. The present CE urine creatinine assay showed a good correlation with HPLC and with Jaffe methods (r = 0.98 and r = 0.97, respectively; p < 0.0001). The uCr in the morning urine samples of 34 healthy males (M), 38 healthy females (F), and 83 male athletes (A) was 10.4+/-6.1 mmol/L, 10.8+/-8.1 mmol/L and 13.2+/-6.5 mmol/L, respectively. The uCr difference (p < 0.02) between M and A and a correlation (p < 0.05) with age in A were observed.