Medical Physics · 1986 · 11 citations · 0 references
MedicineRadiologyMedical PhysicsRadiation DosimetryRadiation PhysicsAbsorbed DoseMonte Carlo DosimetryRadiation SpectroscopyRadiation TherapyX-ray Data AnalysisX-ray SpectroscopyBeam Model ValidationLaplace TransformLaplace Transform AnalysisMonte Carlo Spectrum SimulationRadiation Dose EstimationX-ray Spectral Analysis
We have made a comparative investigation of the estimated spectra obtained by the Laplace transform analysis of the transmitted exposure data measured in an absorption study of a 4-MV x-ray beam. Four transform pair models currently used with this method have been evaluated. It has been determined that the Archer-Wagner model provides a valid representation of the measured transmission data and yields an estimated spectrum which most closely resembles a Monte Carlo spectrum calculated for a 4-MV therapeutic x-ray beam available from a typical medical accelerator.