Translation stage for a scanning x-ray optical interferometer

Peter Becker, Peter Seyfried, H. Siegert

Review of Scientific Instruments · 1987 · 20 citations · 3 references

Concepts

Abstract

A translation stage is described which allows rectilinear displacements of up to 200 μm with a load of 20 N. Lateral deviations from the rectilinear movement do not exceed a few nanometers. Tilts are reduced to a few nanoradians. By means of a piezoelectric drive, slow continuous motion with vibration amplitudes below 0.1 nm is achieved. Residual tilting errors due to machining tolerances and nonideal coupling of the drive are reduced to nrad levels by magnetic compensation elements which are remote controlled.

References

3